WIKA operating instructions digital pressure gauge, model DG-10

11250313.03 09/2012 GB/D/F/E

GB

8. Faults

In the event of any faults, first check whether the digital pressure gauge is mounted correctly, mechanically

and electrically.

8. Faults

Faults

Causes

Measures

No display

Batteries discharged

Batteries inserted the wrong way around

Change batteries

Insert the batteries with the

polarity correct

Display remains on 0, despite

there being pressure

Instrument turned too far

Sensor wire broken

Replace instrument

Constant display signal upon

change in pressure

Mechanical overload caused by overpressure

Replace instrument

Signal span too small

Mechanical overload caused by overpressure

Replace instrument

Signal span varies

EMC interference sources in the environment

Operating temperature too high/low

Remove sources of interference

Observe the permissible

temperatures

Signal span inaccurate

Operating temperature too high/low

Observe the permissible

temperatures

Deviating zero point signal

Medium or ambient temperature too high/low

Observe the permissible

temperatures

If complaint is unjustified, we will charge you the complaint processing fees.

CAUTION!

If faults cannot be eliminated by means of the measures listed above, shut down the digital

pressure gauge immediately, and ensure that pressure and/or signal are no longer present,

and secure the instrument from being put back into operation inadvertently. In this case,

contact the manufacturer. If a return is needed, please follow the instructions given in

chapter 9.2 "Return".