WIKA operating instructions air2guide model A2G-45

9

GB

40389228.01 09/2012 GB/D/F/E/I

6. Commissioning, operation

6. Commissioning, operation

Installation

When connecting the instrument, the process lines must be free of

pressure.

Observe the suitability of the instrument for the media to be

measured

Observe maximum pressures

Electrical connection

The instruments are designed to operate with safety extra-low voltage

(SELV). When wiring up the instruments, the technical specifications

for those instruments should be followed. In the case of a sensor with

transmitter, as a rule, the transmitter should be operated in the middle of

the measuring range, since deviations can occur at the range limits.

The ambient temperature of the transmitter electronics

should be kept constant.

Current/voltage spikes from switching the power supply

on or off must be prevented by the customer.

Current consumption

35 mA + switch (7 mA each) + instrument with automatic zero

adjustment (20 mA)

0 … 10 V output (10 mA), max. 80 mA

5.2 Packaging

Do not remove packaging until just before mounting.

Keep the packaging as it will provide optimum protection during transport

(e.g. change in installation site, sending for repair).

5.3 Storage
Permissible conditions at the place of storage

Storage temperature: -20 ... +70 °C

Protect the instruments from moisture and dust.