Young Sensors Relative Humidity & Temperature Probes Models 41342 User Manual

R. m. young company

Advertising
background image

R. M. YOUNG COMPANY

2801 Aero-Park Drive , Traverse City, Michigan 49686 U.S.A.

TEL (231) 946-3980 FAX (231 ) 946-4772

MODEL 41342

PLATINUM TEMPERATURE PROBE

INSTRUCTION SHEET 41342-90

REV B062309

INTRODUCTION

The Model 41342 Platinum Temperature Probe is an accurate 1000

ohm Platinum RTD temperature sensor mounted in a weatherproof

junction box. The probe is designed for easy installation in YOUNG
Multi-plate and Aspirated Radiation Shields.

INSTALLATION

For accurate measurements, the temperature probe should be installed

in a protective radiation shield. Use of the probe without a radiation

shield may result in large errors due to solar heating. The probe installs

easily in YOUNG naturally ventilated or aspirated shields. For best

performance, the probe and shield should be placed in a location

with good air circulation clear of large masses (buildings, pavement,

solar panels...), exhaust vents, electrical machinery, motors, water

fountains and sprinklers

MAINTENANCE

The temperature probe is designed to offer years of service with minimal

maintenance. If necessary, the probe may be periodically checked or

recalibrated using normal bath calibration methods. NIST traceable

calibration is available from YOUNG at additional cost.
NOTE: The terminal marked "EARTH GND" should be connected

to properly grounded tower or grounding conductor as close to the

sensor as possible. Failure to do so may result in damage due to

static discharge.

CE COMPLIANCE

This product has been tested and shown to comply with European CE

requirements for the EMC Directive. Please note that shielded cable

must be used.

Resistance vs. Temperature for 1000 nominal probe

Temperature Coefficient = 0.00375 / /°C

1226.445

140

1205.659

130

1184.837

120

1163.978

110

1143.081

100

1122.148

90

1101.177

80

1080.169

70

1059.124

60

1038.042

50

1016.922

40

995.766

30

974.572

20

953.340

10

932.069

0

910.759

-10

889.407

-20

868.013

-30

846.576

-40

825.093

-50

Resistance °F

°C Resistance

60

1226.445

50

1189.005

40

1151.445

30

1113.764

20

1075.963

10

1038.042

0

1000.000

-10

961.837

-20

923.550

-30

885.132

-40

846.576

-50

807.873

Transfer function calculated from manufacturer's data:

C° = (1.1279x10

-5

*

R

2

) + (2.3985x10

-1

*

R) - 251.1326

F° = (2.0302x10

-5

*

R

2

) + (4.3174x10

-1

*

R) - 420.0387

WARRANTY

This product is warranted to be free of defects in materials and construc-

tion for a period of 12 months from date of initial purchase. Liability is

limited to repair or replacement of defective item. A copy of the warranty

policy may be obtained from R. M. Young Company.

SPECIFICATIONS

Measuring range:

-50 to +50°C

-50 to +150°F

Accuracy at 0°C:

±0.3°C

±0.1°C (optional)

Time Constant:

42 seconds in 43408 shield.

Sensor type:

1000 Platinum RTD

Output signal:

4 wire RTD

Recommended Cable: 2 pair shielded, 22 AWG (#18723)
Recommended Radiation Shields:

Model 41502 Compact Aspirated Radiation Shield

Model 41003P Multi-Plate Radiation Shield

Declaration of Conformity

R. M. Young Company

2801 Aero Park Drive

Traverse City, MI 49686 USA

Model 41342 PLATINUM TEMPERATURE PROBE

The undersigned hereby declares on behalf of R. M.

Young Company that the above-referenced product, to

which this declaration relates, is in conformity with the

provisions of:

Council Directive 2004/108/EC (December 15, 2004)

on Electromagnetic Compatibility

David Poinsett

R&D Manager

Advertising