Next Advance SP1800 User Manual

Page 32

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1600/1800 Programmable Syringe Pump

Publication #1200-02

27

9/15/04

8.4.4 Example 4: Complex Dispenses with External Synchronization

A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate
during the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the
pump waits for a start trigger before starting the next dispense.

The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set
of dispenses, the syringe is refilled by the amount infused, 17.25 ml. Then the buzzer beeps, to alert the
operator to the start of the first set of dispenses. The process is then repeated.

Phase

Function

Rate

Volume

Direction

1

RATE

750.0 ml/hr

0.5 ml

Infuse

Phase

Function

Rate

Volume

Direction

2

RATE

300.0 ml/hr

1.5 ml

Infuse

Phase

Function

3

BEEP

Phase

Function

4

PS:00

Phase

Function

5

LP:02

Phase

Function

Rate

Volume

Direction

6

RATE

750.0 ml/hr

0.5 ml

Infuse

Phase

Function

Rate

Volume

Direction

7

RATE

300.0 ml/hr

1.5 ml

Infuse

Phase

Function

8

BEEP

Phase

Function

9

LP:ST

Phase

Function

10

PS:60

Phase

Function

Rate

Volume

Direction

11

RATE

500.0 ml/hr

3.75 ml

Infuse

Phase

Function

12

LP:03