32. UPDATE
The function contains two modules which enable updating:
o
Area of a user: SOFTWARE
o
Metrological parameters: WEIGHING MODULE.
Updating process takes place automatically by loading data from a data
storage device connected to balance’s USB port.
Procedure
o
Prepare a data storage device with update file. Required file extension:
*.lab
o
Connect the data storage device to balance’s USB port
o
Select menu <Software>
o
The display previews content saved on the data storage device, search
for the update file and press its name.
o
Updating process takes place automatically.
The process of updating the weighing module takes places analogously to
above procedure, but the required extension of the update file is: *.me01.
Caution:
Weighing module updating procedure is accessible to authorized service points only.
If the update file is incorrect, the balance displays a message on process error –
balance parameters remain unchanged.
33. SYSTEM INFORMATION
This menu contains data on balance specification and installed applications.
Most of parameters in this menu are for the information purpose only.
Parameter <Ambient conditions>, enables previewing present ambient
conditions in a balance: temperature, humidity, pressure (if a balance features
corresponding ambient conditions sensors).
On selecting parameter <Print settings> the balance sends all settings to a
printer.