Printout
Contains type of printout that is related to a working mode. Printout
takes place on pressing PRINT key on balance’s overlay.
Accessible options:
•
STANDARD PRINTOUT
Possible declaration of printout content: HEADER, WEIGHING DATA
and FOOTER.
printout components marked as <YES> in the menu are printed on
pressing printout activating function key.
•
NON-STANDARD PRINTOUT
Database of printouts enables selecting one of available non-
standard printouts visible in menu <PRINTOUTS>, or designing a
unique printout which is automatically added to the databases.
Caution: Means of designing printouts is described in point 15.5 of this
user manual.
−
Air buoyancy compensation
Contains parameters allowing the user to switch on the correction and
enter data relating to density of the sample and to density of the air.
Caution: function operates only for weighing mode.
For instruction and information on air buoyancy compensation operation
see the previous section of this manual.
10.8. Minimum sample
The weighing mode settings comprise a function of <Minimum sample>. The
function is enabled on specifying in the database <Database/Minimum
sample> parameters on methods for determining the minimum sample and
the values of minimal sample weight specific for a method. In standard
version, this database item is not filled in with data.
Activities related to determining minimum sample and adding data to
the database: <Databases/Minimum samples> are accessible only for
an authorized “RADWAG” personnel.
Should balance operator use the minimum sample mode, and data on minimal
sample weight are not entered to balance menu, then please contact the local
representative of “RADWAG” company.
An authorized RADWAG representative determines the minimum sample
dependent on packaging mass. The process is carried out with mass
standards, in balance place of use and in accordance with the requirements of
the accepted quality system. The obtained values are entered to balance
software: <Databases/Minimum samples>.
A minimum sample determining method may have a few tare values defined,
along with assigned values of minimum samples, expiry dates of the carried
out measurements and entered data. The settings are not to be edited by