Start mass adjustment – RADWAG WPW/T 60/HR3/FH Waterproof Control Scales User Manual
Page 92
Advertising

23.2. Start mass adjustment
It is possible to adjust only a start mass, it helps to correct the start zero
when the span does not change.
Procedure:
• Enter
<P10 USER CALIB.> according to 11.2. and then:
• Unload the scale,
• Press
. During adjusting a start mass you will see:
ADJ. START MASS, in the bottom line,
• After completing this procedure the scale will return to the
following window:
• Return to weighing, saving parameters.
Return to weighing:
See 11.3.
- 92 -
Advertising
This manual is related to the following products:
- WPW/T 30/HR3/FH Waterproof Control Scales WPW/T 15/HR3/FH Waterproof Control Scales WPW/T 15/HR2/FH Waterproof Control Scales WPW/T 6/HR2/FH Waterproof Control Scales WPW/T 3/HR2/FH Waterproof Control Scales WPW/T 60/H3/FH Waterproof Control Scales WPW/T 30/H3/FH Waterproof Control Scales WPW/T 15/H3/FH Waterproof Control Scales WPW/T 15/H2/FH Waterproof Control Scales WPW/T 6/H2/FH Waterproof Control Scales WPW/T 3/H1/FH Waterproof Control Scales WPW/T 1,5/H1/FH Waterproof Control Scales WPW 60/H5/K Multifunctional Scales WPW 60/H4/K Multifunctional Scales WPW 60/H3/K Multifunctional Scales WPW 60/C2/K Multifunctional Scales WPW 6/H2/K Multifunctional Scales WPW 6/F1/K Multifunctional Scales WPW 300/H6/K Multifunctional Scales WPW 300/H5/K Multifunctional Scales WPW 300/C3/K Multifunctional Scales WPW 300/C2/K Multifunctional Scales WPW 30/H3/K Multifunctional Scales WPW 30/F1/K Multifunctional Scales WPW 30/C2/K Multifunctional Scales WPW 3/H1/K Multifunctional Scales WPW 150/H6/K Multifunctional Scales WPW 150/H5/K Multifunctional Scales WPW 150/H4/K Multifunctional Scales WPW 150/H3/K Multifunctional Scales WPW 150/C3/K Multifunctional Scales WPW 150/C2/K Multifunctional Scales WPW 15/H3/K Multifunctional Scales WPW 15/H2/K Multifunctional Scales WPW 15/F1/K Multifunctional Scales