Example 2:

? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.


Example 3:

^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.



28.6. Continuous transmission

The indicator can work in a continuous transmission mode. It can be
switched on or off in parameters or using RS232 commands.

Frame format sent by the indicator when continuous transmission is set:

1

2

3 4

-12 13 14 15 16 17 18

stability space

sign

mass space

unit

CR LF

Stability character

[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range

sign

[space] for positive values or
[-] for negative values

mass

9 characters justified to the right

unit

3 characters justified to the left

command

3 characters justified to the left



28.7. Configuring printouts

If some information included are redundant or not sufficient and there is
a necessity of changes you can design a non-standard printout. There is
possible to create up to four standard printouts (see the chapter. 17).