Buffer for weighings, Eighings, Buffer for w – RADWAG PUE C41H Indicator User Manual

Page 40

Advertising
background image

Return to weighing:

See 11.3.



15.1.5.2. Buffer for weighings

Users can declare the quantity of performed measurements to be saved
in the internal buffer in the scale in case of operating OFF-LINE (no
transmission to „E2R SYSTEM”). After reconnecting with „E2R SYSTEM”
all measurements from the internal buffer will be sent to the database
of the computer program.

Procedure:

• Enter

< P3.1 COMPUTER > according to 11.2. and then:

• Type on the scale keyboard the required buffer length and confirm

by pressing button

.

Caution:
The buffer equal to 0 results in saving all weighings in the database
in case of operating OFF-LINE.

- 40 -