29. ERROR MESSAGES
ERROR ”XXX” ESC – RETURN TO PREVIOUS SETTING (where: XXX –
parameter name) – confirmed with ENTER of a wrong value in the user’s
menu,
Err2
- Value beyond the zero range,
Err3
- Value beyond the tare range,
Err4
- Calibration mass or start mass adjustment error
(
±1% for weight, ±10% for start mass),
Err5
- Mass of a single piece lower than the 0.1 of scale division,
Err6
- Mass of all pcs (declared as standard) lower than one
division,
Err8
- Tarring / zeroing operation time exceeded,
NULL
- Zero value from the AD converter,
FULL2
- Measurement
range overflow,
HI
- Display range overflow,
LH
- Start mass error, the mass on the weighing platform
is beyond the acceptable range
±20 of start mass
30. TROUBLE SHOOTING
Problem Cause
Solution
Scale switches
off automatically
„t1” parameter set to
„YES” (Power save)
In <P9 OTHERS> change
<P9.3 AUTO SWITCH.> to „NO”
„LH” appears on
the display
Weight pan not empty
when switching on
Unload the pan. Indication returns
to zero