1 offset / gain register, 2 performing calibrations, 1 offset calibrations – Cirrus Logic CDB5463U User Manual

Page 15: 2 gain calibrations, Cdb5463u

Advertising
background image

CDB5463U

DS805DB2

15

2.5.1

Offset / Gain Register

In the Offset & Gain fields, the offset and gain registers for both channels are displayed in hexadecimal
and decimal formats. These registers can be modified directly by typing the desired value in the hexadec-
imal display fields. There are two types of offset registers: DC offset and AC offset. The AC Offset register
only affects the RMS register values. The DC & AC offset registers are two's complement numbers whose
values ranges from -1 to +1. The gain register value ranges from 0 to 4.

2.5.2

Performing Calibrations

Offset and gain calibrations can be performed on both the voltage and current channels of the CS5463.
Because the initial values in the calibration registers will affect the results of the calibrations, it is generally
a good idea to software reset the CS5463 before running calibrations. A software reset will reset these
registers back to the default values of zero offset and unity gain. Offset calibration should be performed
before gain calibration to ensure accurate results.

2.5.2.1

Offset Calibrations

1. Ground the channel(s) you want to calibrate directly at the channel header(s) J17 and J22 for the voltage

channel, and J24 and J26 for the current channel. The channel(s) could also be grounded directly at the
screw terminals.

2. Press the corresponding AC or DC offset calibrate button (Cal V, Cal I, or Cal Both) beside or below the

corresponding offset registers fields.

The calibration value(s) will automatically update when the calibration is completed.

2.5.2.2

Gain Calibrations

1. Attach an DC or AC calibration signal to the screw terminals, and make sure the corresponding channel

headers (J17, J22, J24, and J26) are set to the input position.

2. Press the corresponding AC or DC gain calibrate button (Cal V, Cal I, or Cal Both) beside or below the Gain

Registers fields.

The calibration value(s) will automatically update when the calibration is completed.

The Calibration Window also contains the Power Offset Register display and adjustment. The user can
read and write the value in the power offset (Poff) register.

Advertising