Chapter 3

23

Operation Verification

Checking the Maximum Leveled Power

3. Ensure that the RF OUTPUT connector is connected to a 50

load.

4. Ensure that the power level entered corresponds to the value listed in

Table 3-1

.

5. Refer to

Table 3-2

for the recommended equipment and measure the output power.

If the measured power level is more than the power level listed in

Table 3-1

, turn the front panel knob

counterclockwise until the measured power level equals the power level in

Table 3-1

.

6. Note the signal generator front-panel amplitude value and use it to repeat the maximum leveled power

check.

If all of the above steps are followed and the check continues to fail, contact the nearest Agilent
Technologies call center for assistance. Refer to

“Contacting Agilent Sales and Service Offices” on

page 16

for contact information.

Table 3-1

Frequency and Power Level Limits

Instrument Option

Amplitude

Standard Option

1EA

Start

Frequency

Stop

Frequency

# Points

E8257D

Option 520

13 dBm

16 dBm

250 kHz

3.2 GHz

50

13 dBm

20 dBm

3.21 GHz

20 GHz

200

E8257D

Option 540

9 dBm

15 dBm

250 kHz

3.2 GHz

50

9 dBm

18 dBm

3.21 GHz

20 GHz

200

9 dBm

14 dBm

20.1 GHz

40 GHz

200

E8257D

Option 550 and

Option 567

5 dBm

14 dBm

250 kHz

3.2 GHz

50

5 dBm

16 dBm

3.21 GHz

20 GHz

200

5 dBm

14 dBm

20.1 kHz

60 GHz

200

5 dBm

12 dBm

60.1 GHz

67 GHz

100

E8267D

Option 520

13 dBm

N/A

250 kHz

3.2 GHz

50

18 dBm

N/A

3.21 GHz

20 GHz

200

E8267D

Option 532 and

Option 544

12 dBm

N/A

250 kHz

3.2 GHz

50

14 dBm

N/A

3.21 kHz

20 GHz

200

12 dBm

N/A

20.1 GHz

40 GHz

200

10 dBm

N/A

40.1 GHz

44 GHz

50