Agilent Technologies E8257D/67D User Manual

Page 35

Advertising
background image

Chapter 3

31

Operation Verification

Checking the Output Power

NOTE

Limit values are due to power meter uncertainty.

Option 550

Option 567

Frequency

Amplitude

Limits

Frequency

Amplitude

Limits

Option

dBm

dB

Option

dBm

dB

60 GHz

None

5

±1.2

60 GHz

None

5

±1.2

1EA

14

±1.2

1EA

14

±1.2

1E1

3

±1.2

1E1

3

±1.2

1E1 & 1EA

10

±1.2

1E1 & 1EA

10

±1.2

67 GHz

None

5

±1.2

67 GHz

None

5

±1.2

1EA

12

±1.2

1EA

12

±1.2

1E1

3

±1.2

1E1

3

±1.2

1E1 & 1EA

10

±1.2

1E1 & 1EA

10

±1.2

Table 3-6

Leveled Output Power Limits, E8267D

E8267D

Option 520

Option 532 & 544

Frequency

Amplitude

Limits

Frequency

Amplitude

Limits

Option

dBm

dB

Option

dBm

dB

250 kHz

None

15

±0.6

250 kHz

None

12

±1.0

3.2 GHz

None

15

±0.8

3.2 GHz

None

12

±1.0

20 GHz

None

18

±0.8

20 GHz

None

14

±1.0

40 GHz

None

12

±1.0

44 GHz

None

10

±0.9

Table 3-5

Leveled Output Power Limits, E8257D cont.