Bacharach AGM300 User Manual

Page 80

Advertising
background image

72

Instruction 3015-4275

Status Register

Register 0x0011h

R/W

10 bytes

Variable Type

Length

Description

Mode

UC

1 byte

Defines Operating Mode of AGM300. 0 = normal Mode; 1
= Zone_Hold Mode; 2 = Diagnostic Mode; 3 = Service
mode. DO NOT MODIFY (use zone hold register or
service mode register to change this parameter)

State

UC

1 byte

Defines AGM300 Current State. 0 = Idle; 1 = Sampling; 2
= Zeroing; 3 = Warm Up, 4 = Pressure Check DO NOT
MODIFY

Measuring

UC

1 byte

Value = 1 if unit is acquiring detector signal for running avg.
DO NOT MODIFY

Active_Zone

UC

1 byte

Current Zone being checked. 0=zone1, 1=zone2, etc.

Max _Alarm

UC

1 byte

Indicates highest non-acknowledged alarm level DO NOT
MODIFY

Alarm_Count

UC

1 byte

Number of alarms that are currently active DO NOT
MODIFY

UNUSED UC

1

byte

UNUSED

Loop_Card

UC

1 byte

Value = 1 if 4-20mA card has been detected DO NOT
MODIFY

Fault

UI

2 bytes

See Note Below

Fault Flag Structure uses bitwise access to 16 bit word as defined below:


Bit 15 (MSB) Clipping Fault

A/D out of range

Bit 14

Zero Fault

Zero voltage outside factory limits

Bit 13

No Flow

No flow on any zone

Bit 12

Purge Flow

No flow on purge

Bit 11

Zone Flow

No flow on a particular zone(s)

Bit 10

Trigger Fault

IR Source clock trigger missing

Bit 9

Unused

Bit 8

Unused

Bit 7

Config. Fault

No Zones Enabled

Bit 6

Unused

Bit 5

Unused

Bit 4

Loop Fault

Open Current loop

Bit 3

RS-485 Fault

Comm Error

Bit 2

Pressure Fault Pressure out of normal operating range

Bit 1

Bench Temp Fault

Bench temperature out of normal operating range

Bit 0

(LSB) Box Temp Fault Box temperature fault