Bacharach AGM300 User Manual

Page 86

Advertising
background image

78

Instruction 3015-4275

Fault Log Register

Register 0x1900, 0x1901h

R

302 bytes

These registers contain the 20 most recent fault events, the time they occurred, and a pointer to the most
recent event. The data is split into 2 registers. The first register contains 200 bytes and the second register
contains 102 bytes. The results of these two register reads should be recombined into the Fault Log Data
Structure after both have been received

Variable Type

Length

Description

Fault

UI

40 bytes

20 most recent fault events. Each event is decoded as
indicated in Fault Flag Structure given after the Status
Register Description

Time

TIM

260 bytes

Time of each fault occurrence. TIM value as defined in
NOTE 2 of Zone Data

Ptr

UC

1 byte

Pointer to most recent event

Unused UC

1

byte



Alarm Log Register

Register 0x1A00h, 0x1A01h, 0x1A02h R 582 bytes

These registers contain the 20 most recent alarm events, the time they occurred, and a pointer to the most
recent event. The data is split into 3 registers and should be recombined into an appropriate structure after all
three registers have been received. Register 0x1A00h contain 200 bytes, Register 0x1A01h contains 200
bytes, and register 0x1A02h contains 181 bytes.

Variable Type

Length

Description

Event

UC

320 bytes

20 most recent alarm events. Each event contains 1 byte
for each zone. Each zone Byte is defined as 0=No Alarm,
1=Leak Alarm, 2=Spill Alarm, 3=Evac Alarm.

Time

TIM

260 bytes

Time of each alarm event. TIM value as defined in NOTE 2
of Zone Data

Ptr

UC

1 byte

Pointer to most recent event

Unused UC

1

byte