Warning – JUN-AIR JAir Dental Compressed Air System User Manual

Page 9

Advertising
background image

9

© 2011, JUN-AIR

We reserve the right to make any alterations which may be due to any technical improvements

Printed in the USA

JUN-AIR Dental Compressed Air System User Guide

CI-101A (Rev B)

CAUTION: Routinely inspect any and
all power cords for cuts and abrasions.
Discontinue use and have authorized
service representative replace cord
if damaged.

ELECTRICAL SAFETY:

• Follow NEC, NFPA 99C and all local codes.
• Qualified personnel must install electrical wiring

and outlets .

• Unit requires a dedicated (separate branch

circuit only) 20 or 30 amp circuit . (See “SITE
REQUIREMENTS”) .

• Never operate unit outside the specified

voltage range 208-253 VAC (see “SITE
REQUIREMENTS”) .

• See “SPECIFICATIONS” for more electrical

information .

• Power indication light on front of electrical cover

displays when system power is on .

ELECTROMAGNETIC INTERFERENCE (EMI):
The JUN-AIR compressor system is designed to
avoid electromagnetic emissions/interference with
surrounding electrical equipment . Due to the vast
assortment of electrical equipment available, it is
possible that some interference may be experienced
by the end customer . If interference
is experienced, the device that is creating
interference should be removed from the room
where the compressor system is located . If the
interference persists, then it may be necessary to
confirm that both devices are connected to isolated
(separated) circuits per “SITE REQUIREMENTS” and
“INSTALLATION - ELECTRICAL” in this manual. If the
problem still occurs, then the two devices should
be moved as far apart as possible . Finally, if the
problem cannot be eliminated, contact an authorized
distributor .

INSTALLATION SAFETY DATA

Install this product

Install this product where it will be weather
protected .

Failure to follow these instructions can result in
death, fire or electrical shock.

Electrical Shock Hazard

WARNING

This product must be properly grounded.
Electrically ground this product per local codes.

Follow all local applied codes prior to installation.

Do not permanently connect this product to wiring
that is not in good condition or is inadequate for
the requirements of this product.

Check the condition of the power supply wiring.

The wire with insulation that is green or green
with yellow stripes is the grounding wire.

in a dry location.

WARNING: Use of an extension
cord is not advisable. An undersized
extension cord will cause a drop in line
voltage and loss of power. Overheating
will result.


WARNING: Electric shock could
occur as a result of improper grounding.
This product must be grounded
according to NEC regulations
and all local codes.

WARNING: Always turn off compressor
and remove power from unit when
servicing or removing electrical cover.
Lock out power at the breaker prior to
servicing.

WARNING: To Avoid Serious Burns:

Do not touch compressor system

during operation.