3power supply, 4preamplifier voltages, 5high voltage – GBS Elektronik MCA166-USB User Manual

Page 18: 6battery

Advertising
background image

MCA166-USB

6.1.3 Power supply

●

Power supply adequate for commonly used NaI, HPGe, CdTe and proportional
counters

●

Current limited mode with indication of overload condition through software; short
circuit protected and automated recovery after overload or shortening

6.1.4 Preamplifier Voltages

●

± 12V; maximum current 60mA for each of the voltages

●

± 24V; maximum current 60mA for each of the voltages

●

Maximum total power at least 2W

6.1.5 High Voltage

●

Single positive or negative modular HV supply reaching from 50V to 3000V

●

Indication of high voltage sign and value by software

●

Calibration within 1% at maximum of bias voltage

●

Maximum current 0.5mA or 0.25W (e.g. 1000V and 0.25mA)

6.1.6 Battery

●

Rechargeable Li-ion battery with no memory effect

●

Deep discharge protected

●

Battery capacity 32Wh at 20°C

●

Charger input: 9V ... 14V / 0.8A

●

Measured values of the battery life time

●

If no detector connected: 30 hours

●

Expected battery life time with detection systems connected:

Standard IAEA NaI detector (ORTEC/SCIONIX): 27 hours

Bias:

500V, 100µA

Preamplifier:

100mW

Miniature CdTe detection probe (SDP310, RITEC, Latvia): 25 hours

Bias:

400V, 100nA

Preamplifier:

+12V/20mA, -12V/9.5mA (@8.4V)

Standard planar HPGe detector (e.g. Canberra, Ortec): 14 hours

Bias:

2000V, 1nA

Preamplifier:

±12V: 35/-25mA; ±24V: 55/-27mA (@8.4V)