Mca166-usb – GBS Elektronik MCA166-USB User Manual

Page 66

Advertising
background image

MCA166-USB

MCA Stabilisation Commands:

Command name

CMD_SET_STABILISATION

Description

The command sets the peak stabilisation.
fl:

0

turns stabilisation off

1

turns stabilisation on

rb+3<fl<re-3

stabilisation to channel fl

rb:

Peak ROI begin (LLD<= begin < end)

re:

Peak ROI end (begin < end <= ULD, (end-begin) < 250)

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

fl

rb

re

end flag

Byte String (HEX)

A5

5A

4D

00

fl

l

fl

h

rb

l

rb

h

re

l

re

h

B9

9B

Used setup value

channels, lld, uld

Affected setup value

stab_state, stab_roi_begin, stab_roi_end

MCA state

STATE_READY

Command name

CMD_SET_STAB_PARAM

Description

The command sets the stabilisation parameters.
st:

time interval (default 10)

sa:

area (default 25000)

Command syntax

Format

integer

integer

integer

long

integer

Parameter

Preamble

command

st

sa

end flag

Byte String (HEX)

A5

5A

67

00

st

l

st

h

sa

l

sa

l

sa

l

sa

h

B9

9B

Used setup value

-

Affected setup value

stab_time, stab_area

MCA state

STATE_READY

66

Advertising
This manual is related to the following products: