GBS Elektronik MCA166-USB User Manual

Page 70

Advertising
background image

MCA166-USB

Command name

CMD_QUERY_USER_DATA

Description

The command reads a 32 bit value from the MCA parameter memory.
e: 0 ... 63 parameter number

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

e

0

end flag

Byte String (HEX)

A5

5A

5E

00

e

00

00

00

00

00

B9

9B

Used setup value

None

Result

see MMCA_QUERY_USER_DATA function in DLL description.

Command name

CMD_SYSTEM_DATA

Description

The command reads the special MCA data.

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

00

End

end flag

Byte String (HEX)

A5

5A

62

00

00

00

00

00

00

00

B9

9B

Used setup value

Result

see MMCA_QUERY_SYSTEM_DATA function in DLL description.

Command name

CMD_QUERY_CENTROID

Description

The command reads the centroid of a ROI (beg<end, LLD<=beg, end<=ULD, end-beg<250).

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

beg

End

end flag

Byte String (HEX)

A5

5A

5F

00

beg

l

beg

h

end

l

end

h

00

00

B9

9B

Used setup value

channels, lld, uld

Result

see MMCA_QUERY_CENTROID function in DLL description.