GBS Elektronik MCA166-USB User Manual

Page 67

Advertising
background image

MCA Detector Power Commands:

Command name

CMD_SET_PREAMPLIFIER_POWER

Description

The command sets the preamplifier power(pp).
0X80

-24V

on

0X40

+24V

on

0X20

-12V

on

0X10

+12V

on

0XF0

all on

0X00

all off

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

pp

0

end flag

Byte String (HEX)

A5

5A

4E

00

pp

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

preamplifier_power

MCA state

STATE_READY

Command name

CMD_SET_BIAS

Description

The command sets the detector high voltage

42

and controls the HV-inhibit-input.

hv:

0 ... 3000

i = 0

Inhibit off

i = 1

”Canberra- or DFG-mode”, HV shut down if inhibit input < 0.5V

i =-1

”Ortec-mode”, HV shut down if inhibit input = +5V

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

hv

I

end flag

Byte String (HEX)

A5

5A

4F

00

hv

l

hv

h

i

l

i

h

00

00

B9

9B

Used setup value

None

Affected setup value

detector_bias, stab_state

MCA state

STATE_READY

42 The high voltages will ramp up or down by the MCA166-USB.

67

Advertising
This manual is related to the following products: