IKA RV 10 control FLEX auto User Manual

Page 49

Advertising
background image

49

Connect the water outlet hose (supplied) by pushing it onto the nipple

until the stop inside the connection socket is reached (fig. 18).

Caution! Ensure that the condenser inlet and condenser outlet are con-

nected correctly.

Automatic distillation cannot be carried out if the water inlet and outlet

hoses are not fitted according to the instructions and illustrations acc.
fig. 14, since this would cause the temperature readings for the inlet and
outlet to be measured incorrectly.

Release the hose connection using the supplied tool.

Water

Connect the water inlet to the water supply (Figure 16a; Figure 16b with

optional RV 10.5001 water regulator valve). Please refer to the technical
specifications relating to the water supply.

The RV 10.5001 water regulator valve is not suitable for use on the

condenser unit because it would reduce the flow rate too greatly

Connect the water hoses to the glass condenser (short hose (O) = lower

outlet, long hose (P) = upper inlet) and fix hoses in place (fig. 17).

Note: For special condensers see “Description of special condensers” section.

Fig. 16a

Fig. 16b

Fig. 17

O

P

Fig. 18

Fig. 19

Inlet
condenser

Outlet
condenser

Water

inlet

Water

inlet

Advertising