Error codes – IKA RV 10 control FLEX auto User Manual

Page 68

Advertising
background image

68

Any malfunctions during operation will be identified by an error message on the display.
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated. The lift can be
operated again.

Proceed as follows in such cases:

Switch off device using the main switch

Carry out corrective measures

Restart device

Error

Effect

Cause

Correction

Communication Distillation cancelled in

IR interface obstructed

- Press ESC to confirm

error bath

"Auto-distillation" mode

Heating bath switched off

- Switch on heating bath

- Check and clean IR interface

No rotation

Distillation stopped,

Rotary drive does not start or

- Switch device off and on at mains switch

no rotary drive

exhibits very large speed variation

No temperature Distillation cancelled in

No differential temperature increase - Press ESC to confirm

difference

"Auto-distillation" mode

of cooling water

- Check flow of cooling water

- Check that the cooling water hoses are fitted correctly

Temperature

Distillation stopped

Differential temperature outside

- Press ESC to confirm

out of range

measurement range (> 7 K)

- Check that the cooling water hoses are fitted correctly

No vacuum

Distillation stopped

Vacuum target value not reached

- Press ESC to confirm

Vacuum cannot be controlled

- Check that the glassware and hose connections

are properly sealed

- Switch on the pump

- Check the solenoid valve connections

No venting

Vacuum cannot be vented

No venting of condenser

- Switch off the device

- Vacuum sensor faulty

- Manually vent the glassware using the stop

valve

Water valve

Water runs continuously into

Water valve not closed

- Press ESC to confirm

not closed

cooling circuit

- Check the connection

Flow rate

Distillation cancelled in

Flow rate of cooling water out of range - Press ESC to confirm

out of range

"Auto-distillation" mode

- Readjust flow rate

- Switch off flow rate control (for dry ice coolers)

Condenser

Distillation stopped

Condenser overload

- Press ESC to confirm

overload

Distillate may be able to enter the

- Change process parameters, e.g.

vacuum pump

- Increase pressure

- Reduce bath temperature

- Increase condenser flow rate within specified range

Distillation

Distillation stopped

The system detected that there was

- Press ESC to confirm

dry

no more medium in the evaporating - Alternatively, activate the “Drying” option

flask during the distillation.

in the “Settings” menu

Lift not in

Distillation stopped

Obstruction in path of lift

- Press ESC to confirm

end position

Lift motor switches off

- Check the path of the lift.

- Remove any obstructions.

Calibration error Calibration reset to factory setting

Out of calibration range

- Press ESC to confirm

(max. +/-0.5 K)

- Check that the sensor is connected correctly

- Carry out a new calibration procedure

Memory error

Calibration parameter for vacuum reset

Error by checking contents of memory Vacuum must be recalibrated

to factory setting

Vacuum leak

Distillation stopped

Leak in vacuum system

Press ESC to confirm

Locate leak

Error codes