IKA RCT basic User Manual

Page 19

Advertising
background image

19

Controlling the Medium temperature

limit via ETS-D or Contact Thermometer

The preferable method for controlling the average temperature is
with the ETS-D or contact thermometer. After the set point tempera-
ture has been adjusted, this results in a short heating-up time, practi-
cally no temperature drift and only minor fluctuation in temperature.
In addition to precise control function, the adjustabel ”MAX-

TEMP” function makes the ETS-D a temperature limiter that

has a direct effecton the average temperature. When this

“MAXTEMP” is exceeded, for example because of a defect

in the control circuit or unintentional of the “Safe Temp” or

set point temperature on the RCT or HCT, the ETS-D imme-

diately and permanently turns off the separate circuit of the

RCT or HCT.

The stirring function then continues to run at the speed that was set
before the malfunktion.
A 6-pin jack is located on the rear side of the instrument for connec-
ting the ETS-D, the contact thermometer or the contact plug. The
electronics of the instruments returns a test current that must flow via
connector pins 3 and 5 for the heating plate to heat up.

Safety contact thermometers

acc. to DIN 12 878 class 2 or acc. Gerstel are connected with a three-
wire cable, the test current flows through the contact thermometer.
Safety function:
If the test current is interrupted because of e.g. breakage of contact
thermometer or falling out of the cable plug, the heating cuts off.
The required plug allocation is indicated on a reference plate above
the diode bushing.
Contact thermometer without safety circuit

acc. to DIN 12 878 class 0. The instrument only heats if the test
current circuit is closed by an electrical connection of the plug pins
3 and 5.

2-wire connecting cables:

Connect plug pins 3 and 5 of the instrument plug.
3-wire connecting cables:

Here the test current circuit can also be produced in the
terminal head of the contact thermometer (connect plug pins
2 and 3).
A 3-wire cable with the required bridge is available (accesso-
ries).

Settings
For the detailed instructions for settings and limit values, please
refer to the operating instructions for the instrument you are
connecting.

The desired average temperature can be adjustedon the

ETS-D or contact thermometer. In addition, a “MAX-

TEMP” can be set for the ETS-D, in other words, a maxi-

mum average temperature can be set for malfunctions in

combination with the RCT or HCT. The required surface

temperature of the heating plate can be selected with

the “Temp” dial.
Adjusting the “Temp” dial to the maximum adjustable tempe-
rature will result in the fastest possible heating time, but the
average temperature may fluctuate to values above the set-point
temperature on the contact thermometer, for example. By
adjusting the “Temp” dial to approximately twice the set-point
value (thus with a set-point ot 60°C the Temp dial would be set
to 120°C), you will reach a good compromise between a fast
heating time and overshooting the set point.
If you adjust the Temp dial to exactly the set-point temperature,
the medium will not reach the set-point temperature, since some
loss of the heat will always occur between the heating plate and
the medium.
The maximum heating plate temperature can be adjusted

will the “Safe Temp” screw.

R_HCT_042011

Advertising