Error codes
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short desciption of the fault.
Error code Cause
Effect
Solution
E3
Temperature inside device too high
Heating off - Switch off device and allow to cool down
E4
Motor blockage
Heating off - Switch off device
Motor off
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in
connector for the motor
E6
Break in safety circuit
Heating off - Plug in contact plug (L)
- Plu in PT 1000 contact thermometer / temperature sensor
- Replace faulty connecting cable, plug, or contact thermometer
E24
Surface temperature (temperature of control sensor):
Heating off - Switch off device until the surface temperature of the base
of the base plate is higher then the setted safe tempera-
plate is lower than the selected safe temperature limit
ture limit
- Set ahigher safe temperature limit
E44
Surface temperature (temperature of safety sensor):
Heating off - Switch off device until the surface temperature of the base
of the base plate is higher then the setted safe tempera-
plate is lower than the selected safe temperature limit
ture limit
- Set ahigher safe temperature limit
E25
Heating and switching element monitoring
Heating off - Switch off device
- Set the safe temperature limit > 55 °C
see also „Fuctional check of inactivating the safety circuit”
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in
connector for the heating element
E26
Difference between temperature of safety sensor and
Heating off - Switch off device
temperature of control senor
- Warning! Only to be carried out by authorised service personnel:
control temperature > (safety temperature + 40 K)
Carry out an internal test on the device to check the plug-in
connector for the temperature sensor
E46
Difference between temperature of safety sensor and
Heating off - Switch off device
temperature of control senor
- Warning! Only to be carried out by authorised service personnel:
safety temperature > (control temperature + 40 K)
Carry out an internal test on the device to check the plug-in
connector for the temperature sensor
Any malfunctions during operation will be identified by an error
message on the display (E and F).
Proceed as follows in such cases:
F
Switch off device using the main switch (A)
F
Carry out corrective measures
F
Restart device
R_HCT_042011