Error codes

If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short desciption of the fault.

Error code Cause

Effect

Solution

E3

Temperature inside device too high

Heating off - Switch off device and allow to cool down

E4

Motor blockage

Heating off - Switch off device

Motor off

- Warning! Only to be carried out by authorised service personnel:

Carry out an internal test on the device to check the plug-in

connector for the motor

E6

Break in safety circuit

Heating off - Plug in contact plug (L)

- Plu in PT 1000 contact thermometer / temperature sensor

- Replace faulty connecting cable, plug, or contact thermometer

E24

Surface temperature (temperature of control sensor):

Heating off - Switch off device until the surface temperature of the base

of the base plate is higher then the setted safe tempera-

plate is lower than the selected safe temperature limit

ture limit

- Set ahigher safe temperature limit

E44

Surface temperature (temperature of safety sensor):

Heating off - Switch off device until the surface temperature of the base

of the base plate is higher then the setted safe tempera-

plate is lower than the selected safe temperature limit

ture limit

- Set ahigher safe temperature limit

E25

Heating and switching element monitoring

Heating off - Switch off device

- Set the safe temperature limit > 55 °C

see also „Fuctional check of inactivating the safety circuit”

- Warning! Only to be carried out by authorised service personnel:

Carry out an internal test on the device to check the plug-in

connector for the heating element

E26

Difference between temperature of safety sensor and

Heating off - Switch off device

temperature of control senor

- Warning! Only to be carried out by authorised service personnel:

control temperature > (safety temperature + 40 K)

Carry out an internal test on the device to check the plug-in

connector for the temperature sensor

E46

Difference between temperature of safety sensor and

Heating off - Switch off device

temperature of control senor

- Warning! Only to be carried out by authorised service personnel:

safety temperature > (control temperature + 40 K)

Carry out an internal test on the device to check the plug-in

connector for the temperature sensor

Any malfunctions during operation will be identified by an error
message on the display (E and F).
Proceed as follows in such cases: