Operation – IKA RET basic User Manual

Page 17

Advertising
background image

17

092008

Commissioning

Stirring

Heating

Connecting

external

thermometers

(direct temper-

ature control in

the media)

REO basic

RET basic safety control

+

Move device switch (A) to the OFF position

+

Insert the mains power cable into the power socket (K)

+

Move device switch (A) to the ON position

Operating mode B is preset

Operating mode A is preset (see “Operating modes”)

Operation

+

Adjust the motor speed using the rotary knob (D)

The value selected will be shown on the display (F)

+

Set the start point for the agitation function by pressing the rotary knob (D)

The displayed value will flash until the desired motor speed is reached

i

Ensure that the protective film is removed from the base plate before use!

+

Move device switch (A) to the OFF position

+

Detach contact plug (L)

+

Attach a DIN 12878 (Class 2) compliant safety contact thermometer or a

PT 1000 temperature sensor to the socket (L)

+

Move device switch (A) to the ON position

Temperature sensor PT 1000

The actual temperature for the temperature sensor shown on

display (E) will correspond to the temperature of the media.
The LED "external temperature sensor" (H) will be lit

Contact thermometer e.g. ETS-D5

Follow the operating instructions for the contact thermometer

The LED "decimal point for the temperature display" (J) will be lit
When a contact thermometer is connected, the display (E)
will only show the set-point temperature that has been set

i

i

+

Set the safe temperature limit (see "Setting the safe temperature limit for RET")

+

Adjust the set-point temperature using the rotary knob (C)

The value selected will be shown on the display (E)

+

Set the temperature control mode (see “Setting the temperature control mode for RET”)

+

Set the start point for the heating function by pressing the rotary knob (C)

The set-point and actual temperatures will be shown alternately on the display (E)

When the heating is switched on, the LED “Heating plate” (G) and the LED “Set-point value” (I)

will be lit.

During agitation and standby operation, the display (E) will show if the base plate temperatu
re is above 50 °C.

HOt

The maximum temperature that can be set for the heating plate is 340 °C

i

i

5000800b_RET_REO_100324:6 28.07.10 15:23 Seite 17