The warranty does not cover worn out parts, nor does it apply to
faults resulting from improper use, insufficient care or maintenan-
ce not carried out in accordance with the instructions in this ope-
rating manual.

REO basic

RET basic safety control

Base plate
Dimensions

mm

ø 135

Material

Stainless steel

Heating
Heating power (-5%/+10%) at 230 V

W

600

Heating power (-5%/+10%) at 115 V

W

600

Heating power (-5%/+10%) at 100 V

W

450

Adjustment and display resolution

K

1

Surface temperature

°C

Ambient temperature - 340

Temperature sensor PT 1000 variation

K

< + (0,15 +0,002xITI)

DIN EN 60751 Kl. A
Temperature variation max. at 100 °C

K

+1,5

Heating plate temperature variation

K

+5

Heating plate control hysteresis,

K

+5

no container, center of heating plate at 100 °C
Control hysteresis,

K

+1

with temperature sensor PT 1000*
with ETS-D5*

°C

+0,5

with ETS-D6*

°C

+0,2

Adjustable safety circuit
Safety temperature limit (adjustable)

°C

50 - 360 (+10)

*

The control accuracy values given were determined using the following set-up: 500 ml water in 600 ml glass beaker, magnetic stirring bar 40, 600 rpm, 50 °C

Subject to technical changes!

23

092008

In accordance with IKA warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact your
local dealer. You may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons for the claim.
You will be liable for freight costs.

Warranty

5000800b_RET_REO_100324:6 28.07.10 15:24 Seite 23

Advertising