Error codes – IKA RET basic User Manual

Page 21

Advertising
background image

21

092008

Any malfunctions during operation will be identified by an error message on the display (E and F).

Proceed as follows in such cases:
+ Switch off device using the main switch (A)
+ Carry out corrective measures
+ Restart device

If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short description of the fault.

Error codes

Error code

Cause

Effect

Solution

E3

Temperature inside device too high

Heating off

- Switch off device and allow to cool down

E4

Motor blockage

Heating off

- Switch off device

Motor off

- Warning! Only to be carried out by authorised service personnel:

Carry out an internal test on the device to check the plug-in
connector for the motor

E6

Break in safety circuit

Heating off

- Plug in contact plug (L)
- Plug in PT 1000 contact thermometer/temperature sensor
- Replace faulty connecting cable, plug, or contact thermometer

E24

Surface temperature (temperature of control sensor): Heating off

- Switch off device until the surface temperature of the base plate

of the base plate is higher than the setted safe tem-

is lower than the selected safe temperature limit

perature limit

- Set a higher safe temperature limit

E44

Surface temperature (temperature of safety sensor):

Heating off

- Switch off device until the surface temperature of the base plate

of the base plate is higher than the setted safe tem-

is lower than the selected safe temperature limit

perature limit

- Set a higher safe temperature limit

E25

Heating and switching element monitoring

Heating off

- Switch off device
- Set the safe temperature limit > 55 °C,

see also “Functional check of inactivating the safety circuit”

- Warning! Only to be carried out by authorised service personnel:

Carry out an internal test on the device to check the plug-in connector.
for the heating element

E26

Difference between temperature of safety sensor and

Heating off

- Switch off device

temperature of control sensor

- Warning! Only to be carried out by authorised service personnel:

control temperature > (safety temperature + 40 K)

Carry out an internal test on the device to check the plug-in
connector for the temperature sensor

E46

Difference between temperature of safety sensor and

Heating off

- Switch off device

temperature of control sensor

- Warning! Only to be carried out by authorised service personnel:

safety temperature > (control temperature + 40 K)

Carry out an internal test on the device to check the plug-in
connector for the temperature sensor

5000800b_RET_REO_100324:6 28.07.10 15:23 Seite 21