Warranty

In accordance with IKA

®

warranty conditions, the

warranty period is 24 months. For claims under
the warranty please contact your local dealer. You
may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons
for the claim. You will be liable for freight costs.

The warranty does not cover worn out parts, nor
does it apply to faults resulting from improper use,
insufficient care or maintenance not carried out in
accordance with the instructions in this operating
manual.

Error codes

The fault is shown by an error code in the display (C) as following if the error occurs, e.g. Er 4.

Proceed as follows in such cases:
F

Switch the device switch (A) off.

F

Remove the stirrer tool and remove the instrument from the assembly.

F

Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.

Error code Cause

Effect

Solution

Er 2

Motor current sensor fault

motor off

- Switch off the instrument

Er 3

Temperature inside instrument
is too high

motor off

- Switch off the instrument and allow it

to cool down

Er 4

Motor blockage or overload

motor off

- Switch off the instrument
- Decrease the load on the motor and

restart again

Er 8

Speed sensor fault or overload motor off

- Switch off the instrument

Er 21

Safety relay fault

motor off

- Switch off the instrument

If the actions described fails to resolve the fault or another error code is displayed then take one of the
following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.