1 rcf display and calculation, Rcf display and calculation, 2 product description – Eppendorf 5424 R Centrifuge User Manual

Page 11: Operating manual

Advertising
background image

11

2 Product description

Centrifuge 5424 / 5424 R — Operating manual

2.4.1

Rcf display and calculation

To determine the g-force (rcf) for a specific adapter, you can calculate per DIN 58 970 using the
following formula:
rcf = 1.118 · 10

-5

· n

2

· r

max

n: revolutions per minute (rpm)
r

max

: max. centrifuging radius in cm

Example
In rotor FA-45-24-11, the 0.5 mL adapter has a maximum radius of 7.3 cm. At 7,000 rpm a
maximum g-force of 4,000 x g is reached.

Use the rpm/rcf key to switch the display of centrifugation speed between speed (rpm) and
g-force (rcf). Ensure that the g-force displayed upon switching is standardized to suit the rotor
FA-45-24-11 without an adapter. When other rotors and adapters are used, you can achieve the
following maximum g-forces (rcf) at the maximum speed (see p. 10):

Rotor

Adapters

Max. centrifugation
radius r

max

[cm]

Max. g-force (rcf)
(5424 120 V, 5424 R)

Max. g-force (rcf)
(5424 230 V/100 V)

FA-45-24-11,
FA-45-24-11-Special

without adapter

8.4

21,130

20,238

0.2 mL

6.3

15,848

15,179

0.4 mL

8.4

21,130

20,238

0.5 mL

7.3

18,363

17,558

0.6 mL

8.4

21,130

20,238

F-45-18-11-Kit

without adapter

7.2

18,111

17,347

0.2 mL

5.1

12,829

12,288

0.4 mL

7.2

18,111

17,347

0.5 mL

6.1

15,345

14,697

0.6 mL

7.2

18,111

17,347

FA-45-32-5-PCR

without adapter

7.4

18,615

17,829

EN

O

perating manual

Advertising
This manual is related to the following products: