1 clean menu and parameters, 2 setting the parameter for clean, 3 execute the clean function – Eppendorf TransferMan4 m User Manual
Page 67: Clean menu and parameters, Setting the parameter for clean, Execute the clean function
Advertising

Operation
TransferMan
®
4m
English (EN)
67
6.14.1
Clean
menu and parameters
Abb. 6-4:Menu – Clean
Fig. 6-4:
Menu –
Clean
6.14.2
Setting the parameter for
Clean
1. Set
Home speed
.
2. Set
Clean distance
.
3. Close the menu.
6.14.3
Execute the
Clean
function
Prerequisites
• Parameters are set for
Clean
.
•
Clean
is assigned to a softkey.
• The capillary is moved out of the medium with the set parameters.
Parameter
Value
Range of values
Increment
Standard
Home speed
Set the speed of the clean
function
5 – 10000
5
7500
Clean distance
Set distance in μm
0 – 20000
5
10000
Set the
Clean distance
above the medium boundary
1. Press
Clean
.
The screen for
Clean
appears.
2. Execute the function with
Clean
.