4 terminate the clean function, 15 piezoxpert function, 1 piezoxpert menu and parameters – Eppendorf TransferMan4 m User Manual
Page 68: Terminate the clean function, Piezoxpert function 6.15.1, Piezoxpert menu and parameters
Advertising

Operation
TransferMan
®
4m
English (EN)
68
6.14.4
Terminate the
Clean
function.
• The function is terminated.
• The application screen appears.
6.15
PiezoXpert
function
6.15.1
PiezoXpert
menu and parameters
Abb. 6-5:Menu – PiezoXpert
Fig. 6-5:
Menu –
PiezoXpert
1. Press
Back manual
.
Parameter
Value
Range of values
Increment
Standard
Synchr.
PiezoXp.
Switch synchronization
between TransferMan 4m
and PiezoXpert on or off
OFF/ON
PiezoXp.
speed
Set the advance rate of the
manipulator in μm/s while
the PiezoXpert triggers a
pulse.
5 – 10000
5
300
PiezoXp. dist.
Set the length of the
movement step of the
manipulator in μm while the
PiezoXpert triggers a pulse.
1 – 2000
1
20