Operation
InjectMan
®
4
English (EN)
66
6.12.2
Executing the
Step injection
function
Prerequisites
• A FemtoJet is connected.
•
Step injection
is assigned to a free softkey.
1. Activate
Step injection.
2. Set all parameters.
3. Set the injection time on the FemtoJet.
4. Close the menu.
• The application screen displays
Step injection
.
5. Trigger the injection with the joystick key.
• The function is executed.
6.13
Home
function
The
Home
function quickly moves the capillary out of the work area and is suitable for a
quick change of capillary.
6.13.1
Home
menu and parameters
Abb. 6-4:Menu – Home
Fig. 6-4:
Menu –
Home
6.13.2
Setting the parameter for
Home
1. Set the speed.
2. Set the offset.
3. Close the menu.
Parameter
Value
Range of values Increment Standard
Home speed
Set the speed of the Home
function
5 – 10000
5
7500
Home offset
Set vertical offset in μm
5 – 20000
5
0