6.15
PiezoXpert
function
6.15.1
PiezoXpert
menu and parameters
Abb. 6-6:Menu – PiezoXpert
Fig. 6-6:
Menu –
PiezoXpert
Parameter
Value
Range of values Increment Standard
Synchr.
PiezoXp.
Switch synchronization
between InjectMan 4 and
PiezoXpert on or off
OFF/ON
PiezoXp. speed
Set the advance rate of the
manipulator in μm/s while
the PiezoXpert triggers a
pulse.
5 – 10000
5
300
PiezoXp. dist.
Set the length of the
movement step of the
manipulator in μm while the
PiezoXpert triggers a pulse.
1 – 2000
1
20
PiezoXp. axial
Axial movement on or off
(corresponds to the set
injection angle)
On = movement in the X and
Y-axis.
Off = movement in the X-axis
ON/OFF
–
OFF