4 installation, Installation – Eppendorf InjectMan NI 2 User Manual

Page 22

Advertising
background image

75

To avoid having to reposition every time a capillary is changed, the
positioning aid can be mounted on the universal capillary holder so it fits
in the centre hole in the A head.

Note

: The connections between the control board, the mains power supply unit and the motor

module must only be plugged and unplugged with the device switched off.

Note

: Do not place the control board down resting on the joystick! When connecting cables,

place the control board on its side on an even surface and secure it against toppling over!

–

The connection to the motor module/injector (e.g. FemtoJet

®

) is made via the

corresponding connectors, as shown in Figure 4.

–

Check the voltage and frequency against the rating plate before connecting the mains
power!

–

Use only the mains power supply unit supplied by Eppendorf!

–

The cable is plugged into the mains power supply unit before being connected to the mains.

Important

:

The main power connection is the last one to be connected.

–

Check that the connectors are firmly attached.

Fig. 4: Underside of control board

1 Connection for optional foot switch (SER2; SubD9, male)
2 Connection for injector (FemtoJet

®

, Transjector 5246) (SER1; SubD9, male)

3 Connection for mains power supply unit

(

Important

:

Connect only after the other components)

4 Connection for motor module (SubD25, female)
5 Serial port (SER0; SubD9, male) for computer (RS 232)

4.4 Connecting cables – Startup

1

2

3

4

5

4 Installation

Installation

4

04_Install_en.fm Seite 75 Donnerstag, 6. Dezember 2012 1:32 13

Advertising