5 alarm functions, 1 deactivating, 2 reactivating – Eppendorf New Brunswick Excella E-24 Shaker User Manual

Page 19: 6 temperature setpoint, 7 temperature offset calibration, Alarm functions 5.5.1, Deactivating, Reactivating, Temperature setpoint, Temperature offset calibration

Advertising
background image

Operation

New Brunswick

Excella

®

E-24 Shaker

English (EN)

19

5.5

Alarm functions

In addition to visual alarm, the Excella E-24 has an audible alarm that is activated at predetermined times.
It can be deactivated by using the MUTE function (see

Deactivating on p. 19).

5.5.1

Deactivating

1. Press the SELECT key until the MUTE indicator illuminates.

2. Press the

or

key to display ON, then press the SELECT key.

5.5.2

Reactivating

To reactivate the audible alarm:

1. Press the SELECT key until the MUTE indicator illuminates.

2. Press the

or

key to display OFF, then press the SELECT key.

5.6

Temperature setpoint

Press the SELECT key until the function °C indicator illuminates. The temperature can be set from 5

°C

above ambient temperature to 60

°C (non-refrigerated units) or from 7

°C to 60

°C (refrigerated units).

Increasing or decreasing the setpoint is accomplished with the

or

key.

During operation, if the temperature of the chamber is more than 1.0

°C higher or lower than the

temperature setpoint, an alarm is triggered. This alarm consists of a flashing °C indicator and audible beep.
The alarm will automatically deactivate as the incubator shaker achieves the set temperature.

5.7

Temperature offset calibration

The temperature probe and the temperature controller are calibrated together at the factory. The
temperature probe measures the temperature of the air at the probe’s location, near the heat exchanger
return vent. The controller uses the probe input to adjust air temperature, up or down, to match the
temperature setpoint.

Depending on various conditions within the chamber (flask placement and size, heat produced by growing
organisms, heat losses due to liquid evaporation from flasks, etc.), the display temperature may differ from
temperatures within the flasks themselves.

Advertising