8 factory calibration, 9 speed calibration, 1 calibration adjustments – Eppendorf New Brunswick Excella E-24 Shaker User Manual

Page 20: Factory calibration, Speed calibration 5.9.1, Calibration adjustments

Advertising
background image

Operation
New Brunswick

Excella

®

E-24 Shaker

English (EN)

20

If you wish to have the temperature display (indicated temperature) match the temperature at a given point
or match the average of a series of points within the chamber (actual Temperature), proceed as follows:

1. Let the unit equilibrate at or near the desired temperature. Record the indicated temperature.

2. Record the actual temperature.

3. Calculate the temperature correction value.

Actual Temperature – Indicated Temperature = Temperature Correction Value

4. Press the SELECT key until the function °C indicator illuminates.

5. Simultaneously press the

or

keys. The display will indicate CAL.

6. Using the

and

keys, enter the Temperature Correction Value calculated in Step 3.

7. Simultaneously press the

and

to save the Temperature Correction Value to memory.

5.8

Factory Calibration

To return to the factory calibration:

1. Press the SELECT key until the function °C indicator illuminates.

2. Simultaneously press the

and

keys. The display will indicate CAL.

3. Using the

or

key, set the Temperature Correction Value to zero.

4. Simultaneously press the

and

keys. The rapid pulsing of the °C indicator will stop.

5.9

Speed calibration

To calibrate the shaking speed:

1. Set the shaker to a speed that can easily be measured. If you are using a strobe, minimum speed should

be 250

rpm.

2. Compare the reading on the display to the measured reading.

5.9.1

Calibration adjustments

If an adjustment is needed:

1. Press the SELECT key until the RPM indicator light illuminates.

2. Press the

and

keys simultaneously. The display will indicate CAL.

3. Press either the

or

keys to change the displayed value to match the measured speed.

4. Press the

and

keys simultaneously to save the adjustment.

5. Turn the shaker off using the on/off switch, then turn it back on.

The °C light will pulse rapidly to indicate it is not operating in the factory default mode.