10 power failure, Power failure – Eppendorf New Brunswick Excella E-24 Shaker User Manual
Page 21
Advertising
Operation
New Brunswick
™
Excella
®
E-24 Shaker
English (EN)
21
5.10
Power failure
In the event of a power failure, the Excella E-24 is equipped with an automatic restart function.
If the shaker was in operation prior to the power interruption, when power is restored the shaker will begin
to operate at its last entered setpoint. The LED display will flash and the audible alarm will sound,
indicating that a power failure has occurred. Press any key to stop the flashing in the display and the
audible alarm.