Eppendorf C-76 Classic Water Bath Shaker User Manual
Page 13

13
New Brunswick
Operating manual
3. Slip two lengths of 3/8 inch (9.5 mm) ID hose with hose clamps over the coil
ends. The hoses should be long enough to suit water or coolant supply and return
(drain) length requirements. Be careful to note or label the operation of each
hose.
4. Allow 1/8 inch (3 mm) to 3/16 inch (5 mm) of hose to extend beyond the hose
clamps and secure in place. Ensure that the clamps do not damage the hose.
5. Feed the two lengths of hose through the grommets on the rear panel and reattach
the rear panel to the unit.
6. Connect and secure the two remaining hose ends to their appropriate connections.
One line must go to an open drain or return to a coolant supply, and the other line
must be adapted to a water or water supply. If tap water will be used as a coolant,
the supply hose should be attached to a control valve to minimize water use.