3 belt replacement, Warning – Eppendorf C-76 Classic Water Bath Shaker User Manual
Page 30

30
M1248-0051
C-76 Classic Water Bath Shaker
Operating manual
7.3
Belt Replacement
To replace the drive belt of the shaker:
1. Allow sufficient time for the water bath to cool, then open the
DRAIN VALVE
and
allow the bath to drain. Close the
DRAIN VALVE
and remove the drain hose if
attached.
2. Remove the
PLATFORM
(if equipped),
DRIVE ARMS
,
KNOBS
and
LOCKWASHERS
.
3. Remove the eight Phillips head screws (see Figure 11) and lift the water bath
assembly from the unit just far enough to allow movement. Do not remove the
water bath assembly entirely from the unit.
Figure 11: Water Bath Removal
WARNING!
Always turn off the shaker and disconnect the power cord from the
power supply before changing the belt.
WARNING!
The following procedure should only be performed by an NBS-
authorized service technician.