Eppendorf AF2200 Plate Reader User Manual
Page 23

Overview
Eppendorf
®
PlateReader AF2200
English (EN)
23
Incubation
Incubation is always done at the heating position to ensure inside the instrument proper temperature
distribution.
Incubation can consist of shaking and waiting steps (up to 2 shaking steps and up to 2 waiting steps are
allowed in any combination).
The
Remaining Wait step
waits until the overall incubation time is over (including shaking and waiting
times).
The incubation program element is typically used to perform shaking and waiting at a certain temperature
for a certain time.
1. Enter the appropriate parameters for incubation:
Incubation time
Enter the total time (min. 5 s)
Actions
Available actions:
Shaking, Wait Time
2 wait and 2 shaking actions are allowed. Select actions by
double-clicking or use the arrow keys.
Organize actions by using the up/down keys.
Remaining Wait Time:
mandatory, cannot be deleted or edited
(duration 3 s)