Defining measurements
Eppendorf
®
PlateReader AF2200
English (EN)
47
5.4
Defining Kinetic Measurements
The following example describes a kinetic measurement of a 96 well plate.
Select the 96 well plate (e.g. Eppendorf 96 Flat Clear UV-VIS) from the
Plate definition
drop-down list,
and select all wells in the
Pattern
program element.
Double-click the Kinetic Cycle program element and define as follows:
•
Cycles/Number of cycles: 50
•
Kinetic Interval (intervals between measurements): Select
Use kinetic interval
and enter: 2 minutes
30 seconds.
Double-click the
Absorbance
program element and define as follows:
•
Wavelength/Measurement: 462 nm
•
Read/Number of reads: 25