Troubleshooting analysis – Autoquip COIL TRANSFER CARS User Manual

Page 33

Advertising
background image

33

PROBLEM

POSSIBLE CAUSE AND SOLUTION

Lift does not raise.

The motor rotation for a 3-phase motor may be

reversed. Reverse only two motor electrical leads.

Check for a line or hose leak.

Check for oil shortage in the reservoir. Add oil as

necessary (See Oil Requirements in the “Routine
Maintenance” section.)

The load may exceed the rating. (See the

“Specifications” section.) Remove the excess load.

The suction screen may be clogged, starving the

pump. Remove and clean the screen. Drain and
replace the oil.

The suction line may be leaking air due to a loose

fitting. Tighten as needed.

The breather holes in the reservoir fill plug may be

clogged. Remove and clean.

The voltage at the motor terminals may be too low to

run the pump with the existing load. Check by
measuring the voltage at the motor terminals, or as
near as possible, while the pump is running under
load. Reading the source voltage or pump-idling
voltage is meaningless. Inadequate or incorrect
wiring can starve the motor when the source voltage
is ample. Correct as necessary.

The "Down" valve may be energized by faulty wiring

or stuck open. Remove the solenoid and check.

The motor may be single phasing. Check wiring,

fuses, etc.

The pump may be seized if motor is humming or

blowing fuses on overload protection devices.
Remove the pump. The pump should be able to be
rotated by hand. Check for cracks in the housing.

TROUBLESHOOTING ANALYSIS

Advertising