Caron 6241 User Manual

Page 11

Advertising
background image

6200 series Operations Manual

Rev D 2/14/2012

Page 11 of 60

Preliminary Cleaning

Your new environmental chamber was thoroughly cleaned prior to leaving the factory.

It is recommended however, to disinfect all interior surfaces with a general purpose

laboratory cleaning agent prior to using the product. After cleaning, dry all interior

components with a sterile cloth as necessary.

Installing the Port Stoppers

The unit has an access port built into each side of the cabinet. The ports are designed

to allow customer access for equipment validation and for installation of other

equipment inside the chamber. These ports should be sealed with the provided rubber

stoppers to allow the chamber to function properly. Install the stoppers provided in the

port on each side of the unit.

Installing the Shelves

Each new environmental chamber includes four or five perforated stainless steel

shelves. Each shelf requires two shelf channels for installation. The left and right shelf

tracks are the same. Prior to installation, take time to consider what the size of the

product being placed in the chamber will be and set the shelf spacing accordingly.

Additional shelving can be purchased through CARON customer service if necessary.

To install the shelf channels insert the rear tab on the shelf channel into the rear wall

on the side wall of the chamber. Then insert the front tab into the front pilaster. Push

the entire shelf channel towards the rear of the unit and snap it down into place.

Each shelf is capable of supporting a uniformly

distributed load of 50 pounds. The maximum

chamber capacity is 500 pounds (stationary). An

optional re-enforced floor is available for heavy loads.

Chamber should be empty when being moved.

Do not have multiple loaded shelves out

simultaneously or the chamber may tip.

Advertising
This manual is related to the following products: