Leveling the Unit

Place a level on the middle shelf of the incubator. Adjust the feet until the unit sits level

left to right and front to back. Even if the unit is level without adjustment, the leveling

feet should still be lowered to avoid the cabinet moving while opening and closing the

outer door & prevent a flat spot from forming on the casters.

Connecting the Drain Line

The chamber drain connection is located in the bottom middle of the back of the

chamber. A 3/8” NPT fitting and tubing are supplied in the unit parts kit. Thread the

fitting into the drain connection and slide the tubing into the drain connection. Pull on

the tubing after installation to make sure it is tight. Route the drain tubing to a local

floor drain.

If a local floor drain is not available, a variety of accessories are available through

CARON customer service. These accessories can also be viewed at

www.caronproducts.com

.

For non-humidified chambers operated above 15°C, a drain is not needed. Install the

3/8” plug into the drain connection.

Connecting the Water Supply

To ensure proper operation, distilled or deionized water is required as a supply on units

that have humidity control. If these water sources are not available contact CARON

customer service.

When using a pressurized water source, failure

to connect the unit to a drain could result in

facility flooding.

The drain line relies on gravity to remove water

from the chamber. The drain line must remain

below the chamber to drain properly. Kinks or

elevations in the drain line above the cabinet

drain will not allow the chamber to drain.