6200 series Operations Manual

Rev D 2/14/2012

Page 14 of 60

Connecting Communications & Analog Outputs

This unit comes standard with features such as RS485 communications and analog

outputs. A set of terminals are provided to connect to RS485 communications and

analog outputs.

Analog Outputs

Analog outputs are either milliamps (0-20mA, 4-20mA) or voltage (0-5V, 1-5V, 0-10V,

0-20V) signal output that represents each of the displayed temperature (and humidity)

values. These options can be used for connection to in-house data acquisition,

recorder, or alarm system.

Factory default settings are as follows:

Parameter

Analog Output

Corresponding Value

Temperature

0 – 5 V

-50 – 100 °C

Humidity

0 – 5 V

0 – 100 %rh

Connect shielded wires to the appropriate signal terminals: I(+) for current (mA) or V(+)

for voltage (DC). For both current and voltage outputs, COM(-) is common terminal.

The controller itself must be programmed for the voltage or current signal and

corresponding scale (see Operations Section)