Specifications – Comark KM340 User Manual

Page 39

Advertising
background image

38

Specifications

Temperature Scale

Celsius or Fahrenheit user-selectable

Measurement Range

-50°C to 1300°C, (-58°F to 1999°F)

Resolution

1°C or 1°F

Accuracy

Accuracy is specified for an operating

temperature of 23°C (73°F), for 1

year,not including thermocouple error

+2, -1°C -50°C to 0°C

+4, -2°F -58°F to +32°F

±(0.2% rdg + 1°C) 0°C to 1100°C

+0.2, -0.7% rdg ±1°C +1100°C to +1300°C

±(0.2% rdg +2°F) +32°F to 1999°F

Temperature Coefficient

including Cold Junction

± 0.01% rdg ±0.04°C per °C change

from +23°C

Reading Rate

2.5 times per second

Input Connector

Accepts standard miniature

thermocouple connectors (flat blades

spaced 7.92mm, centre to centre)

Use with two probes

To avoid measurement errors due to a

continuous electrical circuit when

using a probe in T1 and another in

T2 at least one probe should have an

isolated junction.

Ambient Operating Range

0°C to 50°C (32°F to 122°F)

Storage Temperature

-20°C to 60°C (-4°F to 140°F)

Relative Humidity

0% to 80% (0°C to 35°C) (32°F to

95°F)

0% to 70% (35°C to 50°C) (95°F to

122°F)

Display

3

1

/

2

digit liquid crystal display (LCD)

with maximum reading of 1999

Battery

Standard 9V PP3 battery (NEDA

1604 IEC 6f22)

Battery Life

200 hours typical with ALKALINE

battery

Dimensions (Instrument

147mm(H) x 70mm(W) x 39mm(D)

Advertising