Description

If the hardware handshake cannot be deactivated, then the RS-232 cable can be
used as follows:

3 - STATUS

Status LED

4 - ANALOG OUT

Analog outputs (for logging leakage rate and backing pressure)

Electrical insulation (max. 60 V DC, 25 V AC against GND)

Connection plug arrangement

The analog outputs can be freely configured.

► For other layouts, see installation manual for mass spectrometer module

LDS3000.

Fig: 3

Connection with RS-232 cable (in case hardware handshake cannot be

deactivated)

Color

Status

Meaning

Red

illuminates

Unit not functional or defective

Red

flashes

Not ready for operation, communication to mass
spectrometer module is not available

Cyan

illuminates

Ready for operation; communication to mass
spectrometer module available

Green

Flashes quickly

Boot loader active, ready for software update

Green

Flashes slowly

Data reception on RS232

Yellow

Flashes slowly

Data reception on RS485

Voltage range

0 ... 10 V

Precision

±15 mV offset, additional ±1% from measurement (current
output voltage) as linearity error (at 25 °C)

Resolution

typ. 2.5 mV

Load

> 10 kΩ

Pin

Name

Factory setting mass spectrometer module

1

ANALOG-OUT 1

Leak rate mantissa
1 ... 10 V; linear; in used unit

2

ANALOG-OUT 2

Leak rate exponent
1 ... 10 V; 0.5 V / decade; step function;
1 V = 1 x 10

-12

; in used unit

3

ANALOG-OUT-GND -