Function, Technical data 4.3.1, Mechanical data – INFICON I/O1000 I/O module User Manual
Page 14: Electrical data, Ambient conditions, N (see, Chapter 4.3, 2 function, 3 technical data

14
Description
4.2
Function
The I/O module is a device interface between the MSB box of the mass spectrometer
module LDS3000 and an external controller, for example. The I/O module is
equipped with
• one RS-232 connection
• one RS-485 connection
• one analog input
• two analog outputs
• ten digital inputs
• eight digital outputs
4.3
Technical data
4.3.1
Mechanical data
4.3.2
Electrical data
4.3.3
Ambient conditions
Dimensions (W x H x D)
107.6 mm x 89.7 mm x 76.6 mm
Weight
300 g
Supply voltage
24 V DC
Permissible ambient temperature
(during operation)
10 °C ... 45 °C
Permissible storage temperature
-20 °C ... 60 °C
Max. relative
humidity
< +31 °C
80%
+31 °C to +40 °C
decreasing linearly from 80% ... 50%
> +40 °C
50%
Type of protection
IP 20
Pollution degree
II
Max. altitude above sea level
2000 m