Error messages, Parameter 1-3, Examples 1-3 – INFICON UL1000 Helium leak detector Interface description User Manual

Page 7

Advertising
background image

Interface Description

1-3

iin

s

70

e1

.fm

Op

er

at

in

g M

a

nua

l

(1

00

4)

1.4

Error messages

1.5

Parameter

1.6

Examples

0

ERR_OK

command completed

1

ERR_CMD_START

wrong command start (no "*")

2

ERR_ERR_BLANK

illegal blank

3

ERR_CMD_WORD_1

command word 1 illegal

4

ERR_CMD_WORD_2

command word 2 illegal

5

ERR_CMD_WORD_3

command word 3 illegal

6

ERR_DISABLED

control by RS232 not enabled

7

ERR_ARGUMENT

argument faulty

8

ERR_NO_DATA

no data available

9

ERR_BUFFER_OVERFLOW

buffer overflow

10

ERR_INVALID

command invalid

11

ERR_NO_QUERY

query not allowed

12

ERR_QUERY

only query allowed

13

ERR_NOT_IMPLEMENTED

not implemented

<b>

Boolean:

0 / OFF or 1 / ON respectively

<NRf>

Numeric representation format: integer, real (15.6) or exponent (4.5E-7)

[space] [sign] [ddd] [.] [ddd] [e|E[sign]ddd]

Command

answer

*stat? (CR)

MEAS (CR)

mode

*status? (CR)

MEAS (CR)

mode

*read?

(CR)

2.876E-7

(CR)

leak rate according to
programmed unit

*read:pa*m3/s? (CR)

2.876E-6 (CR)

leak rate in a different unit

*start (CR)

OK (CR)

start messurment

*conf:trig1? (CR)

1.0E-9

retrieve trigger 1

*conf:trig1 2.0E-9 (CR)

OK

set trigger 1