Backlight, User mode, Clock – INFICON 2020ppbPRO Portable Photoionization Detector User Manual

Page 59

Advertising
background image

ISP-074-576-P1D

53

To turn the pump off:

2.

Press the MENU soft key, “Unit Setup”, press SELECT, “Pump”, press
SELECT, “Off”, then press SELECT.

Backlight

The backlight function is used to switch the backlighting on and off when there is
insufficient light to read the display.

To switch the backlighting on and off:

1.

Press the MENU soft key and select “Unit Setup”.

2. Select BACKLIGHT and then press the SELECT soft key. Press the

DOWN ARROW soft key to either turn the backlight on or off.

3. Press SELECT soft key to return to the main display.

To extend the operating life of the battery pack, turn the backlighting off when it is
not required.

User Mode

User mode selects one of the three logging modes. The three logging modes are
Logging Off, Tag and Interval mode.

Logging Off mode displays instantaneous readings only. The display is updated
continuously and readings are not datalogged.

Tag mode displays instantaneous readings. The user has the option to manually
store the instantaneous reading as either a data point without a tag, a background
reading, or a sample reading.

Interval mode displays the instantaneous reading, STEL, TWA and PEAK readings.
In interval mode data is logged at a user selected interval value between 1 second
and 999 seconds.

Clock

The Clock function is used to set both the current date and time.

Entering Numbers with the Soft Keys to Set the
Clock

For all information that you must enter, the left, center and right soft keys
correspond to the up, down, and right arrow. See Figure 14.
The up and down arrows are used to change the character highlighted by the
cursor. The right arrow is used to advance the cursor to the next character on the
right. When the cursor is advanced pa
wraps around to the first character again. To accept the changes, press the DONE
soft key when the checkmark is highlighted.